dc.description9th International Conference on Biomedical Imaging Signal Processing -- OCT 18-20, 2024 -- Hong Kong, PEOPLES R CHINA
dc.description.abstractThis study investigates the possible causes of radioactive leakage occurring during the Tc-99m generation process. For this purpose, hanging film examination and annular film thickness calculations are conducted within the needle of the generation unit using micro flow approximations. Additionally, local and line losses are calculated for each component in the Tc-99m generation setup, and the capillarity effect of the needle is also considered. Further modifications are made to the experimental setup to examine the effects of needle diameter, orifice influence, and an increase in the local loss coefficient. Finally, a two-phase, unsteady computational fluid dynamics study is conducted to gain detailed insights into fluid behavior during the elution process. As a result, it is recommended that a mechanical seal system positioned between the middle needle and the vacuum vial could provide an effective solution to the leakage problem.
dc.description.sponsorshipChinese University of Hong Kong,The Jockey Club School of Public Health and Primary Care
dc.identifier.doi10.1145/3707172.3707184
dc.identifier.endpage83
dc.identifier.isbn9.7984E+12
dc.identifier.scopus2-s2.0-85219531098
dc.identifier.scopusqualityN/A
dc.identifier.startpage78
dc.identifier.urihttps://doi.org./10.1145/3707172.3707184
dc.identifier.urihttps://hdl.handle.net/20.500.12662/10737
dc.identifier.wosWOS:001435181400012
dc.identifier.wosqualityN/A
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.language.isoen
dc.publisherAssoc Computing Machinery
dc.relation.ispartof9Th International Conference on Biomedical Imaging, Signal Processing, Icbsp 2024
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.snmzKA_WoS_20260128
dc.subjectTc-99m generation
dc.subjectTwo-phase flow
dc.subjectLeakage
dc.subjectMicroflow
dc.titleInvestigation of Leakage Problem during Tc-99m Generation Process
dc.typeConference Object

Dosyalar