The Reliability and Validity of DrGoniometer, a Smartphone Application for Measuring Elbow Flexion

Küçük Resim Yok

Tarih

2026

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Routledge Journals, Taylor & Francis Ltd

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

This study evaluated the reliability and validity of the DrGoniometer smartphone application for measuring elbow flexion compared to the Universal Goniometer (UG). Fifty-six healthy participants were assessed by two physiotherapists, with anatomical landmarks marked by an independent examiner. Intra- and inter-rater reliability were calculated using intraclass correlation coefficients (ICC), and measurement error was assessed via the standard error of measurement (SEM) and Bland - Altman analysis. DrGoniometer showed excellent intra-rater reliability (ICC = 0.99; SEM = 0.69 degrees) and higher inter-rater reliability (ICC = 0.87; SEM = 1.28 degrees) than UG (ICC = 0.54; SEM = 3.87 degrees). Cronbach's alpha values were also higher for DrG (0.929) than UG (0.704). Although Bland - Altman plots indicated no systematic bias, they revealed wide limits of agreement, suggesting limited interchangeability between tools. These findings support the DrGoniometer as a reliable tool for elbow flexion assessment, especially in contexts requiring high intra-rater consistency and digital precision.

Açıklama

Anahtar Kelimeler

Elbow flexion, smartphone-based goniometry, reliability, validity

Kaynak

Measurement in Physical Education And Exercise Science

WoS Q Değeri

Q2

Scopus Q Değeri

Q1

Cilt

30

Sayı

1

Künye